n When power on, if you cannot hear a sound from cooling fan, please don’t continue to operate. You 

must shut down power to check or replace the cooling fan.

n During furnace running, please don’t touch furnace to avoid any high temperature burn.

n Before operating the furnace please must place the alumina insulation blockbehind the front door of the furnace

 for good insulation performance.  

n Please don’t open the front door of the furnace above 300℃ to prevent insulation inside furnace from cracking.

n If the furnace is used at first time or not use for long time, please preheat the furnace at 300℃ for at least 2 

hours to remove moisture inside to avoid chamber cracking.

n MoSi2 heating element is only suitable for using in air or inert gas envoriments. Other active gas, such as H2, 

Cl2 and SO2, will damage heating element.

n Please don’t use the furnace at 400 – 700℃ temperature range for long time because MoSi2 heating element will be easy to be oxidized in the temperature 

range.

n MoSi2 heating element is very brittle. Please pay a great attention during moving and handling. Also, please 

avoid rapid heating and cooling to avoid the heating element broken. Max. 10℃/min heating or cooling rate is 

suggested.

n Please check heating element for every three months to see if they are in good connecting condition. If 

connection gets loose, please open case and tighten them properly.

n Please always keep inside clean before operation to avoid contamination to your sample.

n Furnace must be used under following condition:

Temperature: -10 – 75℃; Elevate: < 1000 M, Humunity < 85%; Environment: no vibration and conductible dust, explosive, flammable 

and corrosive gases